Introduction
The CMS6000VET Veterinary
Monitor has abundant monitoring functions and can be used for monitoring of cat,
dog and other animals. User may select different parameter configurations
according to different requirements. It adopts 100-240V~,50/60Hz for power
supply, 8'' TFT color LCD for displaying real-time data and waveform, and up to
8-channel waveform and all monitoring parameters can be displayed
simultaneously. A 48mm thermal printer has equipped with the device. In
addition, it can be connected to the central monitoring system via
wired/wireless network so as to form a network monitoring system.
This
device can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and dual-
channel TEMP, etc. It integrates parameter measurement module, display and
recorder in one device to form a compact and portable equipment.
Functions
Standard parameters: ECG, RESP,
SpO2, PR, NIBP, dual-channel TEMP
1)ECG Heart rate(HR)
ECG
waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration
rate(RR)
Respiration waveform
3)SpO2 Oxygen
saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS),
Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP T1, T2, TD
It has
abundant functions, such as audible and visual alarm, trend data storage and
output, NIBP measurement, alarm event marking and drug concentration
calculation, etc.
Features
1)8'' TFT color LCD,
multi-language interface(Simplified Chinese, Traditional Chinese, English,
French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian,
Kazakhstan, Polish, Czech).
2)Fanless design, quiet, energy-saving and clean,
which reduces the possibility of cross-infection.
3)With standard interface,
OxyCRG, trend chard, big character interface and view bed, convenient to
observe.
4)Waveform, parameter color and location can be set
optionally.
5)Operation mode: keys and buttons.
6)Display up to 8-channel
waveform on one screen.
7)Display 7-lead ECG waveform on one screen, and with
ECG step function.
8)Adopt digital SpO2 technology, anti-motion and
anti-ambient light interference, and measurement can be performed under the
circumstance of weak filling.
9)NIBP measurement mode: Manual/AUTO/STAT,
storage for 4800-group NIBP data.
10)Review for 71 alarm events of all
parameters and 60 arrhythmia alarm events.
11)Drug concentration calculation
and titration table functions.
12)Print trend table data with
one-key.
13)Connect to Central Monitoring System by 3G, WiFi or wired
mode.
14)AC/DC, built-in rechargeable lithium battery achieve uninterrupted
monitoring.
15)Anti-high frequency surgical unit,
defibrillation-proof(special leads are necessary).
Performance
▲ECG
Lead mode:3-lead or
5-lead
Lead selection:I, II, III, aVR, aVL, aVF, V
Waveform:5-lead,
2-channel
3-lead, 1-channel
Gain:2.5mm/mV, 5.0mm/mV,
10mm/mV, 20mm/mV, 40mm/mV
Scan speed:12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range:15~350bpm
Accuracy:±1% or ±1bpm, whichever is
greater
Alarm accuracy:± 2 bpm
Resolution:1 bpm
ST-segment
monitoring:
Measurement and alarm range:
-2.0mV~+2.0mV
Accuracy:-0.8mV~+0.8mV ±0.04mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE,
VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,
PVC, TACHY, BRADY,
MISSED BEATS, PNP, PNC
Pacemaker:yes
▲RESPIRATION
Method:R-F(RA-LL)
Impedance
Respiration rate:
Measurement and alarm
range:0~150rpm
Resolution:1 rpm
Measurement accuracy:±2 rpm
Alarm
accuracy:±3 rpm
Apnea alarm:10~40s
Scan speed:6.25 mm/s, 12.5 mm/s, 25
mm/s
▲NIBP
Method:Oscillometry
Mode:Manual/AUTO/STAT
Measurement
interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960
minutes
Measurement period in STAT mode:5 minutes
Measurement and alarm
range:10 ~ 270 mmHg
Resolution:1 mmHg
Cuff pressure accuracy:±3
mmHg
Measurement accuracy:
Maximal mean error:±5 mmHg
Maximal standard
deviation:8mmHg
Over-pressure protection:
Other mode:297±3 mmHg
Dog
mode:240±3 mmHg
Cat mode: 147±3 mmHg
▲SpO2
Measurement and alarm
range:0 ~ 100%
Resolution:1%
Measurement accuracy:70%~100%: ±2%;
0%~69%: unspecified
▲PR
Measurement and alarm range:30 ~ 250
bpm
Measurement accuracy:±2 bpm or ±2%, whichever is
greater
▲TEMP
Channel:dual-channel
Measurement and alarm range:0 ~
50℃
Resolution:0.1 ℃
Accuracy:±0.1 ℃
▲Power
supply:100-240V~,50/60Hz
▲Safety classification:Class I, type CF
defibrillation-proof applied part
Accessories
1)Digital BD7 veterinary
tongue-clip SpO2 sensor (1M)
2)General digital SpO2 sensor extension cable
(2M)
3)Neonatal capsule single tube repeatable cuff (6-11cm)
4)NIBP
extension tube
5)5-lead ECG lead cable (American-Standard)
6)ECG
veterinary clip (*5)
7)Transvaginal TEMP probe
8)Power adapter
9)Power
cord
10)User Manual
Physical
characteristic
Dimension: 235 mm(L) × 141 mm(W) × 222
mm(H)
Weight: 2.5 Kg