Milwaukee MA911B/2 pH Electrode/Probe/Sensor
Product #: MIL-MA911B-2

The MA911B/2 probe is lab grade, gel-filled pH probe for use with Milwaukee Monitors and Controllers.
 
DESIGN FEATURES
Top features include:
 
DOUBLE JUNCTION
The double junction junction design reduces the risk of clogging.
 
SPHERICAL GLASS TIP
The sperical glass tip provide the largest possible surface area for faster results.
 
RESIN BODY
The resin body is tough to withstand accidental knocks and also protects the glass sensing bulb.
 
EXTENDED CABLE
Offers a 6.6 feet cable for increased flexibility in measuring.
 
USE AND CARE
pH meters usually start performing poorly because of problems with the probe. The two parts of the pH probe that cause problems are the glass sensing bulb and the reference junction.
 
SENSING BULB
The glass bulb loses sensitivity with use and will eventually fail. This is true of all pH probes.
 
REFERENCE JUNCTION
The reference junction is a small hole that allows the meter to compare the sample to a reference. The reading is generated based on the electrical difference between the sample and reference. If the junction hole gets clogged, the pH probe will no longer function.
 
You can extend the life of your pH electrode in the following ways.
 
STORAGE
The pH probe needs to be kept hydrated. Long periods of dry storage will damage the sensitivity of the probe. Allowing the probe to dry out may also result in the junction hole getting clogged.
 
Storing in pH probe storage solution or pH calibration buffers will help address both these issues. Do not store in tap water and DI water. This will damage the sensitivity of the probe.
 
The best way to store the probe is with the probe's cap filled with storage solution or calibiration buffer and the cap tightened to prevent leakage. It is also recommended that to store the electrode upright to further reduce the potential for leakage.
 
CLEANING
Residue from the sample can impact the sensitivity of the sensing bulb and block the junction. This is especially true if the sample has a lot of organic material. We recommend regularly soaking your electrode in cleaning solution.
 
CALIBRATION
The pH sensing bulbs become less sensitive over time so make sure that you are calibrating regularly to keep your PRO pH meter accurate. Also, calibrate after cleaning or a long period of storage.
Our Policies

Payment Information

IF YOU ARE BIDDING MORE THAN ONE ITEM - PLEASE WAIT AND PAY AFTER ALL AUCTIONS HAVE FINISHED. We will be happy to COMBINE SHIPPING on multiple purchases shipped in the same box. Combined Auctions / Store Purchases may not span more than 10 days. Please wait and send payment for all items together. Once you have paid for a purchase we can no longer add additional items to the box.

Domestic and International Buyers

We accept PAYPAL via eBay Checkout. Non-Paying Bidder will be reported after 10-15 days. If you need more time, please stay in touch or let us know you want to cancel and we will send you a cancel auction request.

Shipping Info

We ship UPS orders within 24 hours of cleared payment (Monday - Friday). USPS orders ship within 48 hours of cleared payment. To view shipping cost click on the Shipping & Payments tab and enter the quantity you are interested in. We ship out of our warehouse in Indianapolis, Indiana. All items listed on auction are physically in stock. We do have rare cases where inventory can be misplaced/lost in the warehouse, in these rare cases, we will contact you ASAP and offer your choice of a refund, alternative product, and/or ETA if you wanted to wait for the item.

International Buyers are responsible for any taxes, duties, customs fees or additional charges incurred during shipping

We ship USPS and UPS. We cannot ship gift or sample marked international packages or list a reduced value on the customs invoice. Please note electrical requirements for your countries electric grid. All products we sell, unless noted otherwise are 110v 60hz with a USA plug.

Contact Us

Store hours are 9:30 am - 6 pm Eastern time Monday through Friday. Feel free to call message us with product questions, recommendations and/or order status updates.

Click here to visit our store